8.2 Power Settings

The power limit settings (parameter

OP.Hi

) for the STF models on various supply voltages are as

follows. The figures represent the maximum percentage of time that controlled power is supplied

to the elements. Do not attempt to “improve performance” by setting a value higher than the
correct one from the table. TZF has different power settings for the three zones.

* 5000W model with 4 elements up to late 2000.

** model with 4 elements up to 1995.

STF -/610 were known as CTF before late 1996.

Model

Volts:

110V

115V

120V

200V

208V

220V

230V

240V

380V

400V

415V

49

45

41

60

55

49

45

41

78

72

64

59

54

64

59

54

100

92

82

75

69

59

54

49

74

68

78

71

65

78

71

65

84

77

70

61|78|61 57|72|57 51|65|51 46|59|46 43|54|43

46

43

38

35

32

62

57

51

47

43

51

47

43

76

70

63

57

53

63

57

53

59|45|59 54|41|54 50|37|50

Model

Volts:

200V

208V

220V

230V

240V

380V

400V

415V

42

39

34

31

29

34

31

29

61|78|61 57|72|57 51|65|51 46|59|46 43|54|43 51|65|51 46|59|46 43|54|43

37

34

30

28

25

30

28

25

34

31

28

29

27

28

29

27

74|75|74 69|69|69 61|62|61 56|56|56 52|52|52 61|62|61 56|56|56 52|52|52

TZF 16/610

STF 16/610

2-phase+N

3-phase+N

Single phase

STF 16/610

STF 15/610
TZF 15/610
STF 16/450

CTF 15/--/610 **

STF 15/180
STF 15/450
STF 15/--/450 *
STF 15/610

3-phase delta

TZF 15/610
STF 16/180
STF 16/450

TZF 16/610