Carbolite BLF 18/21 User Manual

Page 11

Advertising
background image


BLF


MF23 – 3.11

11

6.0 F

AULT

A

NALYSIS

7.0

A. Furnace Does Not Heat Up

1.

The HEAT light
is ON

 An ohm meter applied to

the element circuit shows
an open circuit

 A heating element has failed

2.

The HEAT light
is OFF

 The controller shows a

very high temperature
or a code such as S.br

 The thermocouple has broken

or has a wiring fault

 The controller shows a

low temperature

 The door switch may be faulty

or need adjustment

 The thyristor fuse may have

blown

 The power thyristor could be

failing to switch on due to
internal failure, faulty wiring
from the controller, or faulty
controller

 There are no lights

glowing on the controller

 Check the supply fuses and any

fuses in the furnace control
compartment

 The controller may be faulty or

not receiving a supply due to a
faulty switch or a wiring fault

B. Furnace Overheats

1.

The HEAT light
goes OFF with the
instrument switch

 The controller shows a

very high temperature

 The controller is faulty

 The controller shows a

low temperature

 The thermocouple may have

been shorted out or may have
been moved out of the furnace

 The thermocouple may be

mounted the wrong way round

 The controller may be faulty

2.

The HEAT light
does not go off with
the instrument
switch and the fault
persists when a 2A
control fuse is
removed from its
fuse-holder

 The power thyristor has

failed “ON”

 Check for an accidental wiring

fault which could have
overloaded the thyristor.
Isolate the furnace if this fault
persists.