Carbolite 2216 User Manual

Page 4

Advertising
background image


2216 Controller


4

MC11 – 1.04

2.4

Altering the Setpoint

With the display at “home”, showing the measured temperature, press Down  or Up  once to
display the setpoint; press again or hold down to adjust it. The display returns to the measured
temperature when no key is pressed for 0.5 seconds.

2.5

Altering the Ramp Rate

It is possible to limit the rate of heating by setting a ramp rate. Press Scroll  until the legend

SPrr

(SetPoint ramp rate) is displayed. Use Down  or Up  to display and adjust the value.

The ramp rate sets the maximum rate of heating or cooling in degrees per minute. A value of

OFF

cancels the ramp rate, allowing heating and cooling at the maximum rate. When this feature is in
use, there is a “working setpoint” which can be viewed at any time by scrolling to

w.SP

and

pressing  or .

Fig 1 and fig 2 indicate the possible difference between running without and with a ramp-to-
setpoint value (depending on the load and the value used).

2.6

2216 Navigation Diagram

OP.Hi

SPrr

Home

List

Output

List

Access

List

1000

1

000

oP

ACCS

OP

codE

*C

*C

setpoint ramp rate
OFF or value

power limit
setting, if
present

output power,
read only

measured temperature
and setpoint; use

/

to alter setpoint

for factory access to
lists and parameters
not available to the
operator