Model GASG201 Operations Manual

Page 8 of 13

Rev D 01/18/2009

The CO

2

/N

2

gas supply should be 99.5% pure

and should not contain a siphon tube. Gas

pressure to the unit must be regulated to less

than 30 psig. Failure to do so could cause

tubing to burst.

SECTION 2 -- INSTALLATION

The model GASG201 Stand Alone Gas Guard from CARON is designed to allow two

tanks of either CO2 or N2 to be connected to an incubator requiring approximately 15

psig of gas pressure. The unit is designed to automatically switch from the primary tank

to the secondary tank when low gas pressure of approximately 10 psig is detected on

the primary tank. This allows for a continuous supply of CO2 or N2 to an incubator after

the primary tank is empty. In addition, the user is notified of a tank empty scenario via

an audible and visual alarm.

Connecting a CO

2

/N

2

supply

The CO

2

/N

2

gas supplies must be equipped with two stage regulators to ensure that the

incoming gas to the GASG201 is regulated to appropriate levels. The high pressure

stage should have a 0-2000 psig range, and the low pressure gauge should adjust from

0-30 psig. When connecting the gas supplies, adjust each tank output to 20-25 psig. If

the appropriate regulators are not available, contact CARON customer service to

purchase them.

High concentrations of many gases can cause

asphyxiation. The use of monitors and alarms is

recommended for areas where gas can collect.

Model GASG201 is designed to work with N2 or

CO2. Do not connect this unit to a combustible gas

source.