Controlling the Optional Fluorescent Lights (LGHT301 & LGHT305)

The diurnal chamber comes with a fluorescent lighting system used to simulate day

and night testing. There are separate temperature and humidity (optional) set-points

that correspond with the lights on (day) and lights off (night). The set-points can also

be made the same if continuous conditions are needed throughout the light cycles.

The diurnal lighting system consists of separate light banks. Each light bank has two

lamps and is mounted on the underside of a shelf. The banks can be located on the

same or separate shelves. Each light bank has a corresponding switch that enables

that light bank to come on when the timer is on. The 6220 series have three

independent controls while the 6240 series have two. Only the light banks that have

their individual switch turned on will illuminate. The green indicator light above the light

bank switch signifies when each light bank is on.

The timer can be used to trigger an event in a profile to switch between two set-points

(if programmed accordingly).

6220 Series

Individual

Light Bank

switches

Light Power

Switch

Light Bank

Indicators

Light Cycle

Timer

6240 Series