Calibrating the O2 Display

Independent device calibration

A port is available in the center of the inner glass door to allow an air sample to be

drawn from the incubator. If the incubator has just been installed and started up, allow

8-12 hours for the incubator to stabilize before making a calibration adjustment. If the

unit has been in operation, allow a full hour after the display reaches the O2 setpoint

before making any calibration adjustments. After stabilization, open the outer door;

insert the calibration device’s sample tube into the port to take a sample.

1) Press the key from the main menu

2) Press the to select Calibration

3) Press the key to enter Calibration

4) Press the to select O2 Sensor

5) Press the key to enter O2 Sensor

6) Press the to increase / decrease the O2 Sensor value (hold for scroll)

7) Press the key to save the O2 Sensor calibration value

8) Press the key twice to return to the main menu