Direct drive, Maintenance and calibration – COATS 1055 Balancer User Manual

Page 22

Advertising
background image

16 • COATS 950/1025/1050/1055 Wheel Balancer

Direct Drive

Maintenance and

Calibration

Preventative Maintenance

The balancer requires only minor maintenance to

keep the unit operating properly.

1. Keep the display clean and clear. Use a vaporizing

cleaner only. Do not use cleaners or solvents, which
leave oily or filmy residues behind.

2. Keep the adapters, cones, faceplate, threaded

arbor, pressure cup, and hub nut clean. Grease and
dirt buildup will cause premature wear and inaccurate
balancing. Clean at least once a day with a vaporizing
solvent.

3. Clean the weight tray, accessory post, and pegs.

Weights stored in a dirty tray will pick up grease and
dirt which may alter their weight or keep them from
securely attaching to the wheel. Use a vaporizing sol-
vent to clean the tray, pegs, and post.

4. Keep the area around the balancer clear. Remove

any tools or other items that are leaning against the
balancer. Keep the area under the balancer clear as
well. Remove any items that may cause the balancer
to not sit level.

5. Use only COATS accessories. Accessories from

other manufactures may not fit or function properly,
and may damage the balancer.

Calibration

The balancer utilizes a "self calibration" scheme. To

keep your balancer in proper calibration, this calibra-
tion procedure should be performed once a month, or
whenever the accuracy of the balancer is questioned.

1. Mount a 14, 15, or 16-inch diameter x 6 - 6 1/2-

inch wide steel wheel assembly on the balancer. It
should be suitable for mounting using the back cone
method and have a properly inflated tire.

2. Enter A, W, and D wheel measurements.

Note: The A entry must be between 5.0 and 8.0, and

W entry must be between 5.5 and 7.0. If the wheel
you have mounted does not meet these require-
ments, select a different wheel and try again.

3. Remove any weights attached to the wheel.

4. Press and hold SHIFT and press 1.

5. Lower the hood and press START.

6. Raise the hood, rotate the wheel until the outer

weight positioning LED flashes, and attach a 4-ounce
calibration weight if in ounce mode or a 100-gram cal-
ibration weight if in gram mode at top-dead-center.

Note: It is critical that this weight be placed accu-

rately to achieve proper calibration.

7. Lower the hood and press START.

8. Raise the hood and remove the weight.

The balancer is now calibrated. If a mistake is made

during this procedure, turn the balancer off, then back
on, and start the procedure over again.

Advertising
This manual is related to the following products: