12

4 .5

Battery Replacement

Unscrew the two screws at the base of the instrument

and slide off the battery cover. Replace the battery with

a 9V PP3 type battery. Replace the battery cover taking

care not to over tighten the two screws.

4 .6

Use with Pressure Test Kit

The C9500 series instruments can be used together

with the optional TK2 pressure test kit as a low cost

calibration system for checking or calibrating other

pressure instruments.

To use TK2 as a calibrator, connect a calibrated C9500

series instrument, suitable for the maximum pressure

to be used, to one of the ports on the hand-pump.

Connect the instrument to be checked or calibrated to

the other port. Read the instructions supplied with the

TK2 kit for details of the pump operation. Ensure the

C9500 meter is zeroed with the pressure relief valve

open before measurements or adjustments are made.

Note: With the C9551 meter it is possible to over-

range the instrument with just one squeeze of the

pump handle. If this should occur unscrew the pressure

release valve slowly until the display again shows a

pressure reading then tighten. The volume adjustment

on the pump may now be used for fine control of the

applied pressure.

4.7

Silicone Version

Although C9500 instruments are generally used for gas/

air measurement, special versions are available with

silicone protection to allow liquid pressure measurement

without sensor corrosion.