CONDUCTIVITY PROBE INSTRUCTIONS

SPECIFICATIONS

Glass diameter:

0.472 inches (12 mm)

Glass length:

4.75 inches (120 mm)

Weight:

3.1 ounces (88 grams)

Cable length:

72 inches (183 cm)

Range:

0.05 micromhos to 200,000

micromhos

Constant K factor:

1

Temperature

compensation:

Automatic with

internal thermistor

Solution contact

construction:

Glass and Platinum (guard —

acrylic with rubber O-rings)

FEATURES

• Small cell sample volume, only 30 milliliters

are required, allow analysis of semimicro

samples and expensive specimens.

• Probe designed so that the platinum elec-

trodes may be easily accessed for cleaning,

examination or replating.

• Probe with a plastic guard to help prevent

breakage.

• Close tolerance cell construction to control

the K factor.

• Small glass mass cell designed to reduce

electrical leakage paths and eliminate extra-

neous readings.

• Probe with a computer-designed shape to

eliminate error-causing air bubbles.

RESULTS

All platinum and glass construction providing vastly

superior results when compared to plastic probes.

Plastic probes leach contaminants/mold released

into samples and develop micro fractures which

trap materials and contribute to cross sample

contamination. Plastic probes also entrap air

bubbles, obscure viewing the sensing element and

require a greater sample volume.

OPERATION

Do not use the acrylic guard for critical analysis,

research analysis, or when measuring extremely

pure water. Remove the guard by sliding it down

and off the glass probe. (A new probe may require a

moderate amount of force to slide it off.) Use the

guard for routine quality control, when critical analy-

sis is not required, and for storage. The guard

reduces the sensitivity of the probe by limiting the

electrical field measurement.

BE CONSISTENT

If you standardize using a known calibration solu-

tion with the guard on, then measurements must be

made with the guard on.

If you standardize with the guard off, then measure-

ments must be made with the guard off.

CLEANUP AND STORAGE

Always rinse the probe after using with deionized/

distilled water and replace the acrylic guard before

storing. Sample solutions that are allowed to dry on

the probe will eventually block out active sites on the

platinum electrodes.

CONDUCTIVITY PROBE STORAGE SOLUTION

Redi-Stor

TM

is the ideal solution for storing con-

ductivity probes. It preserves probe’s cleanliness, it

eliminates growths found when storing in only water,

and it maintains the probe for immediate use with no

conditioning. Cat. No. 4170 Redi-Stor

TM

Probe

Storage Solution 16-ounce bottle.

WARRANTY OR SERVICE

For warranty or service contact:

CONTROL COMPANY

4455 Rex Road Friendswood, Texas 77546 USA

Ph. 281-482-1714 Fax 281-482-9448

E-mail [email protected]

www.control3.com

Control Company is ISO 9001 Quality-Certified by DNV and ISO 17025 accredited as a Calibration Laboratory by A2LA

Cat. No. 4079

Conductivity Probe Instructions

92-4079-00

Rev. 1 071504

© 2004