Traceable wide range thermometer instructions – Control Company 4007 WIDE RANGE THERMOMETER User Manual

Page 2

Advertising
background image

RS-232 PC INTERFACE
The meter features an RS-232 output which allows
data to be exported to a computer or data logger
in real-time via the accessory data acquisition
software. (See the “Accessories” section.)
DISPLAY MESSAGES
“BAT” appearing on the display indicates that the
battery is low and needs replacement (see the
“Battery Replacement” section)
“---” appearing on the display indicates that no
probe is plugged into the meter, or the probe is
damaged.
ALL OPERATIONAL DIFFICULTIES
If this meter does not function properly for any
reason, please replace the battery with a new,
high-quality battery (see the “Battery Replacement”
section). Low battery power can occasionally
cause any number of “apparent” operational
diffi culties. Replacing the battery with a new fresh
battery will solve most diffi culties.
BATTERY REPLACEMENT
An erratic display, faint display, no display, or
“BAT” appearing on the display are all indicators
that the battery needs replacement. To replace the
battery, remove the battery cover screw (located
on the back of the meter). Remove the battery
cover by sliding it in the direction of the arrow.
Remove the exhausted battery and replace it with
a new 9-volt alkaline battery. Replace the battery
cover. Replace the battery cover screw and
tighten securely.

ACCESSORIES

Cat. No. 4237 Data Acquisition System
Accessory
Powerful and easy to use computer data capture/
data logging program works with Traceable

®

Instru-

ments with computer output. Records interval read-
ings from 1 to 10,000 seconds; displays minimum/
maximum readings; and utilizes an alarm mode
that permits the user to be notifi ed visually, audibly,
and by email when an alarm is triggered. Data is
stored to a fi le that can be printed in any report or
spreadsheet format. Networking server/client capa-
bility allows the captured data to be monitored on a
remote workstation and/or by email. It is designed
to work with Windows

®

98/Me/NT/2000/XP/Vista.

Includes a CD, a 6-foot cable (supplied USB and
serial connections) that plugs into the instrument
and computer. Accessory extension cables expand
cable length to 300 feet.
Cat. No. 4008 Surface Probe—
Temperature range is –73 to 760°C. Dimensions:
5-inch probe length; 0.5-inch tip diamter; 8½-inch
overall length. Supplied with 36-inch cable.
Cat. No. 4014 Stainless Steel Probe—
Stainless-steel, triple purpose (liquids, air/gas,
and semi-solids), Type-K probe. Temperature
range is –50 to 700°C. Dimensions: 0.13-inch
diameter; 6¼-inch stem length; 9¾-inch overall
length. Supplied with 40-inch cable.
Cat. No. 4028 Beaded Probe—
Fast-Response, Type-K thermocouple, beaded
probe. Tefl on cable can withstand temperatures
of -40 to 250 °C continuous or 300 °C short-term
use. Dimensions: 0.06-inch diameter probe with
cable length of 4 feet for use in liquids, air/gas, and
semi solids.
Cat. No. 8039 Low-Temperature Probe—
Stainless-steel; triple purpose (liquids, air/gas, and
semi-solids), Type-K probe. Temperature range is
-240 to 220 °C. Dimensions: diameter 0.17 inch;
stem length 12 inches; overall length 17 inches.
Supplied with 8-foot cable.
Cat. No. 8613 High-Temperature Probe—
Ten-foot-long, 0.19-inch diameter braided metal
wire cable with smooth tip measures –73 to 982 °C
continuous or 1093 °C short-term use. For use in
liquids, air/gas, and semi-solids.

TRACEABLE

WIDE RANGE

THERMOMETER

INSTRUCTIONS

®

Control Company Cat. No. 4007

Traceable

®

is a registered trademark of

Control Company.

©

2010

Control Company. 92-4007-00 Rev. 3 12072010

WARRANTY, SERVICE, OR RECALIBRATION
For warranty, service, or recalibration, contact:

CONTROL COMPANY

4455 Rex Road

Friendswood, Texas 77546 USA

Ph. 281 482-1714 • Fax 281 482-9448

E-mail [email protected] • www.control3.com

Control Company is ISO 9001 Quality-Certifi ed by DNV

and ISO 17025 accredited as a Calibration Laboratory

by A2LA.

Advertising