PVX Precision Visual Thickness Gauge

59

Adjusting the Threshold using the Hot Menus

1) Press the MEAS key once to activate measure menu items. Press the

MEAS key multiple times to move right and the ESC key multiple times to

move left, until the THR (threshold) cell is highlighted.

2) Press the UP, DOWN, LEFT, and RIGHT arrow keys to scroll the highlighted

value.

3) Alternatively, press the ENTER key to display the Digits Edit Box.

4) Press the UP and DOWN arrow keys to scroll the highlighted value.

5) Press the LEFT and RIGHT arrow keys to scroll the digit locations.

6) Repeat steps 4 & 5 until the THR value is correctly displayed.

7) Press the OK key to return to the measurement screen, or ESC to cancel

entering the THR.


The user can also access and adjust the Threshold from the tabbed menus.
However, this method is more tedious than making the adjustments using the Hot
Menus
. The procedure using the tabbed menus is outlined below: