Dakota Ultrasonics DFX-8 plus MANUAL2 User Manual

Page 105

Advertising
background image

DFX-8

Ultrasonic Thickness Gauge

101

9.3 Zero Coating

In order to account for very slight electronic differences in transducers of the same
type, frequency, and diameter, the DFX-8 has been equipped with a “zero coating”
feature. This enables the DFX-8 to obtain very accurate readings on coatings,
eliminating potential errors incurred from slight differences in the manufacturing
processes. The procedure is outlined below:

Performing a Coating Zero

1) Press the MULTI MODE key once to activate the measurement mode

options.

2) Use the

UP and DOWN arrow keys multiple times using the left button

wheel to scroll through the sub menu items until Coating Only (CT) mode is

highlighted.

3) Press the MEAS key to select the measurement mode and return to the

measurement screen.

4) Apply a drop of couplant on the transducer and place the transducer in

steady contact with the probe zero disk (battery cover) and obtain a steady

reading.

Note: The coating measurement displayed will potentially be a value greater or

less than 0.

5) Press the MENU key once to activate the menu items tab. Press the MENU

key multiple times to tab right and the ESC key multiple times to tab left until

Advertising