Profile mode, Auto start mode – Despatch RAD1-42-2E with Protocol Plus User Manual

Page 31

Advertising
background image

23

Profile Mode

1.

Press the Select key until Profile is displayed. “None” may be displayed if a
profile has not been selected or no profiles entered.

2. Press the

key to display the desired profile to run.

3. To start Profile Mode, press the Run key.

The display will change from Stop to Run and the segment time remaining,
Temperature Setpoint, Profile #, along with the current segment number, will be
displayed. To return to Stopped Mode, press the Stop key.


Pressing the Run/Hold key while the Profile Mode is in operation will put the control in
Hold status. Press the Run/Hold key again to continue the mode. The Profile LED will
flash to indicate the hold status.

To advance to the next segment while running a profile, press the Select and UP arrow
keys at the same time.

If Link To is set to Standby in the Program Page, at the End of Program/Profile,

1. Cycle Complete LED indication goes ON.
2. Controller beeps if End of Cycle beep is enabled.
3. Heater/control output keeps controlling oven temperature at last Soak setpoint.
4. All events programmed (if relay cards installed and programmed as an event) for

the last Soak Segment stays active.


Note that ramping down too fast may cause the high limit relay to trip unexpectedly if
the high limit band feature is used. This can be avoided by using a separate cooling
profile that does not utilized the high limit band and then jumping to that profile to
perform rapid cooling.

Auto Start Mode


The Auto Start Mode allows the control to start Manual, Timer, or Profile mode
automatically at a preset time and day. See the Auto Start Page in Setup Mode for the
time, day, and operating mode settings.

To activate the Auto Start Mode,

1. On Auto Start page, Enable is set to Yes.

2. LCD reads Active on line 1 in Auto Start Mode.

3. On Auto Start page Enable set to No, will deactivate Auto Start Mode.

Advertising