Det-Tronics PIRECL Infrared Hydrocarbon Gas Detector PointWatch Eclipse User Manual

Page 28

Advertising
background image

26

95-8526

14.2

oRdERing infoRmation

pointwatCh eClipSe DeteCtor
When ordering, please refer to the PIRECL Model

Matrix.

Calibration eQUipMent
Pointwatch Eclipse calibration kits consist of a sturdy

carrying case containing two 3.6 cubic foot (103 liter)

cylinders of specified gas, a regulator and pressure

indicator, three feet of tubing, barbed nozzle for direct

application to the device, and a calibration wind shield to

contain the gas in high wind applications.

Methane, 50% LFL, 2.5% by volume

006468-001

Ethylene, 50% LFL, 1.35% by volume

006468-003

Propane, 50% LFL, 1.1% by volume

006468-004

Butane, 50% LFL, 0.8% by volume

006468-006

PointWatch Eclipse Regulator

162552-002

Eclipse Calibration Bag

006682-002

Other calibration gases are available. Consult the

factory for specific information.

Spare partS
Weather Baffle with Inlet Nozzle,

007165-002

with Hydrophobic Filter

Weather Baffle with Inlet Nozzle,

007165-001

without Hydrophobic Filter

Weather Baffle w 1/16” NPT cal

007165-004

gas inlet, with Hydrophobic Filter

Weather Baffle w 1/16” NPT cal

007165-003

gas inlet, without Hydrophobic Filter

Calibration Port Cover

009192-001

Calibration

Magnet

102740-002

Silicone Free Grease

005003-001

O-Ring, 3.75” i.d., for wiring

107427-040

compartment cover

O-Ring, 3.25” i.d., for front

107427-053

flange (internal)

O-Ring, 2.44” i.d., for weather baffle

107427-052

aSSiStanCe
For assistance in ordering a system to meet the needs of

a specific application, contact:

Detector Electronics Corporation

6901 West 110th Street

Minneapolis, Minnesota 55438 USA

Operator: (952) 941-5665 or (800) 765-FIRE

Customer Service: (952) 946-6491

Fax: (952) 829-8750

Web site: www.det-tronics.com

E-mail: [email protected]

table 7—troubleshooting Guide

fault Condition

Corrective action

Low 24 volts

24 vdc operating voltage is out of range. Verify proper wiring to the detector and correct

voltage output from the power source. Power supply faults are self-clearing when the

condition is corrected. If the fault does not clear, consult the factory.

Dirty Optics

Perform cleaning procedure, then recalibrate as required. (Refer to “Maintenance” for

details.)

Calibration Fault

If the calibration process is allowed to time-out, the fault is set and can only be reset

with a successful calibration. Check the gas bottle to ensure that there is enough gas to

complete the calibration. Are conditions too windy for a successful calibration? If so, use

a PointWatch Eclipse Calibration Bag (P/N 006682-002). Always calibrate with a Det-

Tronics calibration kit for Eclipse with correct regulator. Be sure that the calibration gas

being used matches the configured setting. If the fault is still present, perform cleaning

procedure, then recalibrate.

Negative Signal Output

This fault is indicated when the signal output drops below –3% LFL. Normally detection

capability is not compromised in this condition. The device was probably zero calibrated

with background gas present. If the condition persists, purge with clean air and repeat the

zero calibration.

Calibration line active at start-up

The only way to clear this fault is to correct the wiring and reapply power. Be sure that the

calibration line is not shorted and that the calibration switch is open. If the fault does not clear,

consult the factory.

Other Faults

Consult the factory.

Advertising