Unpacking your hotblock, Installation requirements, Electrical requirements – Environmental Express HotBlock CURRENT User Manual

Page 8

Advertising
background image

800.745.8218 / 843.576.1147 / www.environmentalexpress.com

Environmental Express

• 6

HotBlock™:

Operation and Instruction Manual

Unpacking and Installation

Unpacking Your HotBlock™:

1. Remove the HotBlock™ from the shipping container by lifting from the

bottom of the block. The lid should not be used for lifting.

2. Your HotBlock™ is shipped with metal screws securing the bottom panel.

The metal screws must be removed before operating your HotBlock™.
Remove the metal screws and replace them with the PVC screws and
rubber feet included with your shipment. The corrosion-resistant PVC
screws and rubber feet are designed to secure the bottom plate.

Important: Do not over tighten the PVC screws!

Installation Requirements

Locate the HotBlock™ under a fume hood with a minimum face velocity of
100fpm, and allow a minimum of 2" of space on all sides. The following
environmental conditions should be observed:

• Ambient temperature range:

5-30°C

• Ambient relative humidity:

0-90%RH

• Altitude:

sea level to 2500 meters

HotBlocks™ are rated as Pollution Degree 2 and Installation Category 2.

Electrical Requirements

Required Voltage:

120 volts, ~60Hz, 15A

(all HotBlocks™ are also available in 240V with CE mark)

Power should not vary greater than±10%. Use the supplied heat-resistant power
cord or equivalent to connect to the power supply.
For safety reasons, a separate power receptacle should be provided for each unit in
the system. Do not use extension cords or outlet adaptors. Make certain that power
outlets are earth-grounded at the grounding pin.
See individual specifications for each HotBlock™ model, pages 3-4.