Troubleshooting

Troubleshooting

Problem

Possible Cause

Solution

Temperature Controller Display Blank
Red Lighted Rocker Switch not lit

Unit turned off

Turn on Red Lighted Rocker Switch

Unit unplugged or no power to unit

Plug unit into 120 VAC/20A outlet

Main fuse blown

Replace 15A/250V fuse

Internal failure

Call for service

Red Lighted Rocker Switch lit

Units with S/N 2056 and below

High Temp Switch tripped

Reset High Temp Switch

Green Pilot Light not lit

High Temp Switch tripped

Reset High Temp Switch

Units with S/N 2057 and above

heater turned off

Turn on Black Rocker Switch

Internal failure

Call for service

Unit does not heat to operating
temperature

Green Pilot Light lit

See above

Process Temperature not set

Set correct operating temperature

High Temp Switch tripped

Reset High Temp Switch

Units with s/n 2057 and above

heater turned off

Turn on Black Rocker Switch

High Temp Switch failure

Call for service

High Temp Switch trips

Block temperature above

Set operating temperature to

repeatedly

safe range

190°C or below

High Temp Switch failure

Call for service

Block temperature exceeds set
value by more than 1°C

Internal failure

Call for service

Temperature Controller
displays error message

Refer to Controller Manual