Specifications – Braun bp1600 User Manual

Page 11

Advertising
background image

11

Specifications

Model:

BP1650 / BP1600

Operation principle:

Oscillometric method

Display:

Liquid Crystal Display

Measuring range:

0–320 mmHg (cuff pressure)
40–199 / min (pulse rate)

Blood pressure
measurement:

20 mmHg (minimum diastolic value)
300 mmHg (maximum systolic value)

Laboratory accuracy:

± 3 mmHg (cuff pressure)
± 5 % of reading (pulse rate)

Problem

Reason

Solution

Readings from
repeated meas-
urements differ
considerably.

Blood pressure is a
fluctuating value. For
healthy adults, deviations
of 10 to 20 mmHg are
possible.

none

Cuff is not properly
fitted.

Ensure that the monitor is fitted as
follows:
- on the palm side of your wrist
- approx. 2 cm away from the

palm line.

Readings were not
consistently taken at
heart level.

For each measurement, hold the
monitor at heart level.

Talking, coughing,
laughing, moving etc.
when measuring will
influence the reading.

While measuring, relax, keep still,
do not move or speak.