Chapter 5: Voltage-to-Temperature

In standard TCDA modules the low temperature is -50

°

C and the

high temperature is 200

°

C. These temperatures correspond to an

output voltage between 0 and 5VDC. The Voltage-to-Temperature
Conversion Equation for standard TCDA modules is shown below.

NOTE: Do not use this equation if you have a modified TCDA
module.

The following values are used for standard TCDA modules:

-50

°

C < Temperature < 200

°

C

0VDC

<

Vout

<

5VDC

Vout is measured on DB-25 pin 8 ( pin 9 for channel 2)

resolution

=

20mV/

°

C = 0.02V/

°

C

Voff

=

-1.0VDC


TCDA modules with modified temperature measurement ranges

will be supplied with temperature range, values for Voff, and
resolution, and a modified Voltage-to-Temperature Conversion
Equation.

The Fahrenheit-to-Centigrade and Centigrade-to-Fahrenheit
Conversion Equations are shown below.

Voltage-to-Temperature Conversion Equation

Temperature C

Vout

Voff

resolution

° =

+

C-to-F and F-to-C Conversion Equations

(

)

C

F

= ×

5

9

32

F

C

= Ч +

9

5

32