Basic safety instructions

6

Hei-VAP Value, Hei-VAP Advantage

EN

EN

2.4 Use in explosion-prone areas

Do not use the equipment in explosion-prone areas. The equipment is not protected
against explosions. It is not equipped with Ex or ATEX protection.

2.5 Responsibilities of the operator

 Only operate the equipment in perfect condition, contact a certified Heidolph Service

representative if equipment does not meet the outlined conditions for operation.

 Verify that only qualified personnel operate the equipment with the appropriate per-

sonal protective protection.

 Verify that personnel have received safety instructions for working responsibly and

safely in the laboratory.

 Verify that the rotary evaporator is set up in a safe and suitable location.
 Verify that the basic device is only plugged to or unplugged from the heating bath if

the main switch is switched off and/or the main power supply is disconnected.

Depending on the media (chemicals) used:

 Verify that the set up and the operation are only within the designated buildings with

the proper equipment for laboratories.

 Verify that the rotary evaporator is only operated in conjunction with a ventilation

system (refer to DIN EN 14175 and DIN 12924).

Ventilation system:

at least 10 times the air exchange

monitored for malfunctions

2.6 Responsibilities of the operating personnel

 Verify that the distillation material can be evaporated safely and that the distillation

residue is not explosive.

 Verify that work using an open flame is not conducted in the surroundings of the

rotary evaporator (explosion risk).

 Verify that the flow rate of < 1 m/s is guaranteed when suctioning liquids that contain

combustible materials (electrostatic charge; ignition risk).

 Verify that gases of explosion group IIC do not occur in materials or chemical reac-

tions, such as hydrogen.

 Verify that equipment that is an emission or radiation source (electromagnetic

waves) for the frequency range (3*

10

11

Hz to 3*

10

15

Hz) is not operated or attached.

 Verify that equipment that is an emission or radiation source for ionizing waves or

that is within the ultrasound range is not operated or attached.

 Verify that adiabatic compression and shock waves do not occur (shock wave igni-

tion).

 Verify that the use of substances with the risk of an uncontrolled release of energy

associated with a rise in pressure is prohibited (exothermic reaction; self-ignition of
dust).

 Verify that the glass surfaces are wiped off by using only damp cloths.
 Wear suitable protective clothing for activities on the rotary evaporator a (with gog-

gles and gloves, if necessary).

 Avoid pressure on the LCD screen, it is not intended to be a touch screen interface.
 Verify that the maximum excess pressure of the coolant equals no more than 1 bar

in the condenser.

 Verify that the basic device is only plugged to or unplugged from the heating bath if

the main switch is switched off and/or the main power supply is disconnected.