Configuration

Adjustment

1013 mbar

RS-232...

Sensors...

Display...

Autostart

Off

Defaults

Cancel

- - - - - - - - Back - - - - - - - -

In the menu ”Configuration” the device parameters are preselected.

Adjustment: Adjustment of the pressure transducer under vacuum and/or at atmospheric pressure,

see also section ”Readjustment the controller”.

Adjustment at atmospheric pressure is carried out at an absolute pressure value between 1060 - 700

mbar and under vacuum at an absolute vacuum value between 0 - 20 mbar. In the range between 20

and 700 mbar no adjustment is possible; ---- mbar is displayed.

RS-232: Configuration of the interface, setting of parameters and commands, see section ”Inter-

face”.

Baud rate can be set to 19200, 9600, 4800 or 2400, parity on ”8-N-1”, ”7-O-1” or ”7-E-1”, Handshake

on ”no”, ”Xon-Xoff” or ”RTS-CTS” and remote on ”On” or ”Off”.

Sensors: Selection of the pressure transducer to be controlled (maximum 4 external sensors).

Display: Selection of the device parameters ”Brightness” between 0 - 100%, ”Contrast” between

0 - 100%, ”Sound” ”On” or ”Off”, ”Units” ”mbar”, ”hPa” or ”Torr”, ”Language” ”German”, ”English”,

”French”, ”Italian”, ”Spanish”, ”Turkish”, ”Korean”, ”Chinese”, ”Portuguese”, ”Russian”, ”Polish”,

”Dutch”, ”Japanese”, “Finnish”.

Autostart: If ”Autostart” is set to ”On” the controller restarts a running process automatically after a

mains failure. If this is unwanted, set Autostart to ”Off”.

Attention: If Autostart” is preselected, the process starts immediately after power failure without

pressing any further key. It is the user’s responsibility to ensure that no dangerous status of the sys-

tem due to the automatic start-up can occur and to provide appropriate safety measures. If neces-

sary, the user has to check

prior to starting the process if the option ”Autostart” is enabled.

Defaults: If ”Defaults” is set to ”Load”, the controller is reset to factory set values. All stored programs

and parameters are deleted.

The screen-shot shows the factory-set values.

Preselections

Use the selection knob to select the parameters.

5.8 Function Configuration