INFICON Helium Leak Detector Modul1000 Interface description User Manual

Page 7

Advertising
background image

ASCII Protocol

1-5

jin

s

80

e.f

m

Op

er

at

in

g M

a

nua

l

(1

01

1)

*STATus

status of Modul1000 (INIT, ACCL, STBY, VENT,
WAIT_EVAC, EVAC, MEAS, CAL, ERROR)

[:CAL]

mode calibration
(IDLE, EVAC, OPEN, TUNE, TUNE_RES, CLOSE,
STABLE_CLOSE, WAIT_OK)

[:CALHist 1]
[:CALHist 2]

[:CALHist 12]

CAL History 1 to 12 (date, time, type of calibration,
mode, calibration factor)

[:ADDCALHist 1]
[:ADDCALHist 2]

[:ADDCALHist 12]

additional information for CAL_History entry 1 to 12
(mass, filament, ion current when test leak open, ion
curren when test leak closed, nominal test leak rate)

[:CALMode]

kind of calibration INT_AUTO, INT_MAN,
EXT_AUTO, EXT_MAN, ZERO_POINT

[:ERRor]

current number of error / warning

[:ERRHist 1]
[:ERRHist 2]

[:ERRHist 12]

error history entry 1 to 12.

[:ERRorHist]

error History (1…12)

[:ZERO]

zero

[:RANGE]

measuring mode ULTRA (VACUUM), FINE (SNIFF),
NONE

[:RESULT]

status of Auto leak test

[:TestLog 1]
[:TestLog 2]

[:TestLog 12]

test protocol (for auto test leak)

[:SECINEMEAS]

time since change of measuring mode [s]

[:VALVE]

status of internal valves

[:EXT_VALVE]

status of external valves

[:PREAMPRESistor]

currently used resistance of pre-amplifier 13M, 470M,
15G, 500G

PURGe

purge / gas ballast

*READ

leak rate in chosen unit

[:ATM*CC/S]

current leak rat in atm*cc/s

[:G/A]

current leak rate in g/a

[:MBAR*L/S]

current leak rate in mbar*l/s

[:OZ/YR]

current leak rate in oz/yr

[:PA*M3/S]

current leak rate in Pa*m3/s

[:PPM]

current leak rate in ppm

[:TORR*L/S]

current leak rate in Torr*l/s

*STArt

start

Advertising