Load default values

The "Load default values" button serves for loading of one of the four factory-
provided parameter sets

Default setting: Parameter set 1

6.12.6.7 Main menu

Settings

Monitoring

Calibration Request

It can be selected whether the operator is reminded of the fact that a calibration may
has become necessary or not. The default setting is "Off".

If the calibration request is switched on, a corresponding message will appear when
30 minutes have elapsed after power on or if the temperature of the Modul1000 has
changed by more than 5°C since the last calibration.

Paging Function

If the Modul1000 has a wireless remote control RC1000WL, an audio signal can be
enabled which sounds on the remote control to find it and to identify it.

Default
Parameter set 1

Default
Parameter set 2

Default
Parameter set 3

Default
Parameter set 4

PLC IN Pin 3

START

START_STOP

START

START/STOP

PLC IN Pin 4

STOP

GAS BALLAST

STOP

CAL

PLC IN Pin 5

ZERO

ZERO

ZERO

ZERO ON

PLC IN Pin 6

CAL

CAL

INV GAS BALLAST ON

PLC IN Pin 7

CAL INTERN

CLEAR

GAS BALLAST OFF

PLC IN Pin 8

CAL EXTERN

NOT USED

NOT USED

PLC IN Pin 9

CLEAR

NOT USED

NOT USED

PLC IN Pin 10

GAS BALLAST

NOT USED

NOT USED

PLC OUT Pin 3

TRIGGER 1

Error

MEASURE

EMISSION ON

PLC OUT Pin 4

TRIGGER 2

WARNING

Error

INV TRIGGER1

PLC OUT Pin 5

TRIGGER 3

EMISSION ON

MEASURE

INV TRIGGER2

PLC OUT Pin 6

ZERO ACTIVE

CAL ACTIVE

EMISSION ON

INV ERROR

PLC OUT Pin 7

EMISSION ON

GAS BALLAST

GAS BALLAST

PLC OUT Pin 8

Error

CAL REQUEST

TRIGGER 1

PLC OUT Pin 9

CAL ACTIVE

MEASURE

TRIGGER 2

PLC OUT Pin 10

CAL REQUEST

ZERO ACTIVE

TRIGGER 3

PLC OUT Pin 11

OPEN

TRIGGER 1

OPEN

PLC OUT Pin 12

OPEN

TRIGGER 2

OPEN

PLC OUT Pin 13

OPEN

TRIGGER 3

OPEN

PLC OUT Pin 14

OPEN

REC STROBE

OPEN

RECORDER Pin 2

LR MANTISSA

LR MANTISSA

LR MANTISSA

RECORDER Pin 3

LR EXPONENT

LR EXPONENT

LR EXPONENT

LR-UNIT

mbarl/s

Pa m³/s

Pa m³/s

Pa m³/s

Pressure Unit

mbar

Pa

Pa

Pa

Control location

LOCAL

LOCAL, PLC

LOCAL, PLC

LOCAL, PLC