7.2.1 Starting AutoTune from the Manual Tune Screen on the Laptop
Computer
1
Double-click on the ER IQ icon. See
Figure 7-2 ER IQ Icon
2
Double-click on the Tune icon. See
HINT: Only the Advanced Access level can run Manual Tune. See
8.8.1, Changing Access Levels, on page 8-18
.
Figure 7-3 Manual Tune Icon
3
Wait until the EM and Emission buttons in the Control Panel turn green. Click
the Short Tune icon. See
.
Figure 7-4 Short Tune Icon
CAUTION
Adjusting other parameters without proper training may
damage the instrument.
4
Allow the Short AutoTune to run to completion and click OK. Close the Manual
Tune after receiving the message Final Results are on the LCD. See
.
Figure 7-5 AutoTune Complete