3 adjusting the crystal holder retainer spring – INFICON UHV Bakeable Sensor User Manual

Page 41

Advertising
background image

4 - 7

PN

07

4-

15

4N

UHV Bakeable Sensor Operating Manual

4.3.3 Adjusting the Crystal Holder Retainer Spring

Occasionally, the ceramic retainer may not be secured in the crystal holder. To alter
the retainer retention force, use the following procedure.
Tools required

Scribe or other pointed tool

Needle nose pliers (two required)

Procedure

1

Position the crystal holder with the crystal aperture oriented downward.

2

Insert the point of the scribe between the inside edge of the crystal holder cavity
and one of the two wire segments that protrude into the crystal cavity (see

Figure 4-5

(a)).

Figure 4-5 Location of the transition point

3

Using the scribe, gently remove the spring from its groove in the
crystal holder cavity.

4

Refer to

Figure 4-5

(b) to determine the direction in which the ‘transition point’

must be relocated, to attain the desired retention forces. Moving this transition
point approximately 1.59 mm (1/16 in.) is generally sufficient.

5

Grasp the spring, with the pliers, just below the transition point. Use the second
set of pliers to bend the spring as illustrated by the dashed line in

Figure 4-5

(c)

to remove the existing transition point.

6

Use both pliers to form a new transition point according to

Figure 4-5

(b),

thus returning the spring to a shape similar to the solid line delineation
of

Figure 4-5

(c).

Wire

Scribe

Crystal Holder

Move location of transition point in this
direction to decrease retainer retention force

Move location of transition point in this
direction to increase retainer retention force

Location of Transition Point

(a)

(b)

(c)

Advertising