5 update source parameter – INFICON IC6 Thin Film Deposition Controller User Manual

Page 183

Advertising
background image

10 - 13

PN

07

4-

50

5-

P1

F

IC6 Operating Manual

10.4.5 Update Source Parameter

Command = UC <Command ID> <Source number> <Parameter Value>

Command ID = <Byte> See

Table 10-5

.

Source Number = <Byte> Source number 1 - 6

Parameter Value = <Integer> | <Float>

Type of value depends on command ID, see Data Type column in

Table 10-5

.

Response = None (Just header and trailer) .

Table 10-5 Source parameters

QC UC

Cmd ID

Name

(Units)

Range:

Low Limit/

High Limit/

Default

Data Type

Notes

Update Restrictions

and Display Notes

1

(0x01)

DAC Output

0/ 12/ #

Integer

0 = not defined.

Default for source 1 is 1,

source 2 is 2, etc.

Update not allowed if

source is running. Not

allowed to set if the DAC is

set as a recorder output.

2

(0x02)

(Source) Shutter

Output

0/ 38/ 0

Integer

0 = not defined.Update not allowed if source is

running.The same output can be used for multiple

shutters. If multiple shutters have the same output, the

name will list on the I/O all source numbers assigned to

that output.

3

(0x03)

Number of Crucibles

0/ 5/ 0

Encode

0 = 1 1 = 4 2 = 8 3

= 16 4 = 32 5 = 64

Update not allowed if

source is running.

4

(0x04)

Crucible Output

0/ 38 /0

Integer

0 = not defined

Gray font if Number of

Crucibles = 0

5

(0x05)

Turret Feedback

0/ 1/ 0

Encode

0 = No 1 = Yes

Gray font if Number of

Crucibles = 0

6

(0x06)

Turret Input

0/ 28/ 0

Integer

0 = not defined

Gray font if Turret Feedback =
No or if Number of Crucibles =

1

7

(0x07)

Turret Delay

(Seconds)

2/ 180/ 5

Integer

Time out interval with

feedback.Delay interval

without feedback.

Gray font if Number of

Crucibles = 0 or if Number of

Crucibles = 1

Advertising