3 status indicators – INFICON BPG402-Sx ATM to Ultra-High Vacuum Gauge User Manual

Page 35

Advertising
background image

tina46e1-a (2010-03) BPG402.om

35

The admissible data rate depends on a number of factors such as system pa-
rameters and cable length (

→ [4], [6]). It can be set on the gauge or pro-

grammed via DeviceNet.

By means of the "RATE" switch, the data rate can be set to 125
("1"), 250 ("2") or 500 kBaud ("5").

Default data rate setting is 500 kBaud.

If the switch is in any of the "P" positions, the data rate is pro-
grammable via DeviceNet (

→ [1]).


Two LEDs on the gauge inform on the gauge status and the current DeviceNet
status.

LED Description

Dark no

supply

Red/green,

flashing

self test

Green normal

operation

Red

non recoverable error

LED Description

Dark

gauge not online:

− self test not yet concluded
− no

supply,

→ "STATUS MOD" LED

Green,

flashing

gauge online but no communication:

− self test concluded but no communication to other nodes

established

− gauge not assigned to any master

Green

gauge online; necessary connections established

Red, flashing

one or several input / output connections in "time out" status

Red

communication error. The gauge has detected an error that im-
pedes communication via the network (e.g. two identical node ad-
dresses (MAC IC) or "Bus-off")

The gauge is connected to the DeviceNet system via the 5-pin DeviceNet con-
nector (

→ 21).



4.10.2.3 Data Rate Setting

4.10.3 Status

Indicators

"STATUS MOD"
(gauge status):

"STATUS NET"
(network status):

Electrical connections