tira93e1 (2014-04) BPG402-SE.cp

27

Writing the corresponding password (vendor specified value) enables Access
Rights (Index F940 Subindex 03)

Index

SI

DataType

NV Access PM Name

F940 0x01

UDINT RO Measurement Principle

0x02 BYTE

RO

Number of Sensors

0x03

USINT

RO

Access

Rights

0x04 V_STRING

x

RO

SW Version VPG PCB

F9F0 V_STRING

RO Manufacturer

Serial

Number

F9F1 0x01

UDINT RO CDP

Functional

Generation Number
Module 1

0x02

RO

CDP

Functional

Generation Number
Module 2

F9F2 0x01

UDINT RO SDP

Functional

Generation Number
Module 1

0x02

UDINT

RO

SDP

Functional

Generation Number
Module 2

F9F3 V_STRING

RO Vendor

Name

F9F4 0x01

V_STRING

RO Semiconductor

SDP

Device Name Module 1

0x02

V_STRING

RO

Semiconductor

SDP

Device Name Module 2

F9F5 0x01

USINT RW Output Identifier Module 1

0x02

USINT

RW

Output Identifier Module 2

F9F6

UDINT

RO

Time Since Power On

F9F7

UDINT

x

RO

Time Total Powered

F9F8 UDINT RO Firmware

Update

Functional Generation
Number

Measurement principle assigned to the object instance.
The most significant nibble of the parameter represents the sensor type of the first
Module, the second most significant nibble of the parameter represents the sensor
type of the second Module, and so forth.

Measurement Principle (Index F940)

1
2
3
4
5

Capacitance Manometer
Piezo
Heat Transfer
Cold Cathode
Hot Cathode

Number of Sensors (Index F940): In this device are 2 sensors implemented.

Subindex 0x02