2 configuration capacitance diaphragm, 3 configuration trip point 1 – INFICON PSG55x ATM to Medium Vacuum Gauge (EtherCAT) User Manual

Page 24

Advertising
background image

24

tira85e1 (2013-06) PxG55x.cp

Index

SI

DataType

NV Access PM Name

8001

0x01 REAL

x

RW

Zero Adjust Mode

The CDG Sensor is adjusted automatically by the Pirani sensor, if this index is set.


The High Trip Point is calculated by:

High Trip Point = Value

*)

× Percentage

**)

*)

Value from Piezo Sensor (I: 0x6010, SI: 0x11)

**)

Percentage

High Trip Source (I: 0x800E/F, SI: 0x13)

In this case, the High Trip Point value is automatically updated by the device itself.
Any user-written value to the parameter High Trip will be refused with an abort
code.

The Low Trip Point is a fixed value:

The value defined in Low Trip Point Limit is compared with the pressure value
referenced by the Source Index parameter.

High Trip Point

Time t

in-

active

active

in-

active

active

Low Trip Point

Measurement signal

(Pressure p)

Low Trip
Hysteresis

High Trip
Hysteresis

in-

active

Index

SI

DataType

NV Access PM Name

800E

0x01 BOOL

x

RW

High Trip Enable

0x02 BOOL

x

RW

Low Trip Enable

0x11 REAL

x

RW

High Trip Point Limit

0x12 UDINT

x

RW

High Trip Source Index

0x13 REAL

x

RW

Percentage High Trip
Source

0x14 REAL

x

RW

Low Trip Point Limit

0x15 UDINT

x

RW

Low Trip Source Index

0x16 REAL

x

RW

Percentage Low Trip
Source

0x17 REAL

x

RW

High Trip Hysteresis

0x18 REAL

x

RW

Low Trip Hysteresis

0x1A

UDINT

x

RW

Source

Index

4.7.2 Configuration

Capacitance Diaphragm

Subindex 0x01

4.7.3 Configuration Trip

Point 1