tina83e1-a (2014-02)

47

Fasten the measuring chamber (3) by means of the

hexagon socket set screw (4).

MPG50x gauge only: Perform an ATM and HV adjustment

of the Pirani measuring circuit via the <ADJ> button
(

→  39).

A recalibration of the MAG50x gauge is not

necessary.

We recommend to perform a leak test (leak rate

<1×10

-9

mbar l/s) and a function test of the gauge on the

teak detector.

WARNING

WARNING: electric arcing
Helium may cause electric arcing with detrimental
effects on the electronics of the product.
Before performing any tightness tests put the pro-
duct out of operation and remove the electronics
unit.