JULABO FK31-SL Calibration Baths User Manual

Page 26

Advertising
background image

Safety installations, warning functions

26



complete shutdown of the heater and pump is effected.

The alarm is indicated by optical and audible signals (continuous tone).
The following error message appears on the VFD COMFORT-DISPLAY:

Press

for help on the LCD

Depending on the setting of >ControlType - internal or external< the
actual temperature values for both sensors are shown/indicated on both
displays.

Examples: I 94.06

and Ext. : xx.xx°C

Type:

>Warning< or >Alarm<

For the two menu items >OverTemp< and >SubTemp< choose between
a warning message being signaled or a complete shutdown of the main
functional elements such as heater and circulating pump being effected.

OverTemperature: SubTemperature:

If for a sensitive temperature application task adherence to a working
temperature value >Setpoint< is to be supervised, then set high and low
temperature warning values.
In the example below, the >Setpoint< of 85 °C is surrounded by the
values >OverTemp< 87 °C and >SubTemp< 83 °C. The electronics
immediately registers when the actual temperature attains a
temperature out of the limits and it follows a reaction according to what
is set in the menu item >-Type<.

OverTemp 87.0°C

SubTemp 83.0°C

Setpoint3 85.0°C

The warning functions are only triggered when the actual bath
temperature, after start from the „OFF“ or „rOFF“ mode, lies within the
set limits for 3 seconds.