MENU ATC - Absolut Temperature Calibration

ATC serves to compensate a temperature difference that might occur
between circulator and a defined measuring point in the bath tank
because of physical properties.



Example:
1-point calibration

°C

°C

T 1

T


3-point calibration

°C

°C

T 1

T

T

T

1 = Original curve

Principle

:

For ATC calibration, in steady state the bath temperature at the location
of the temperature sensor (CT) is determined at the respective adjusted
working temperature. This value is then set on the circulator in the menu
>ATCalibration< under menu item > CTEMP X <.
This can be a 1-point, 2-point or 3-point calibration.

°C

TT

M

B (INT)

T

°C

TT

M

B (EXT)

T

Pt100

ext


M = Temperature measuring instrument with temperature sensor
B = Bath tank (INTernal or EXTernal)
T = circulator

CT = Temperature on measuring point