Over and Sub temperature warning function

Over temperature

Sub temperature

If the observance of a working temperature value >SETP< has to be
supervised for a sensitive temperature application, then set over and sub
temperature warning values. In the example below the SETPOINT 85 °C is
surrounded by the values OVERTMP 87 °C and SUBTEMP 83 °C. The
electronics immediately register if the actual temperature breaches one of
the set limit values. The resulting reaction is defined in a further menu item.
(See chapter 8.2.

)

1. Press the key

.

2. By pressing the

or

key select the menu >OVERTMP< or

>SUBTEMP<.

1.

Press the

key. The integer digits flash

2.

Change the values to 87. °C and/or 83. °C by pressing the

and

key and confirm with the

key.

The decimal digits flash and can be adjusted if desired.

Confirm once more by pressing the

key.

See above examples.

• The warning functions are only activated if the actual bath temperature

remains within the set limit values for 3 seconds after switch-on.

Recommendation:

Set the over temperature warning value >OVERTMP< 5 °C to 10 °C above
the working temperature setpoint.

Set the sub temperature warning value >SUBTMP< 5 °C to 10 °C below the
working temperature setpoint.