JULABO LH 85 High Dynamic Temperature System User Manual

Page 13

Advertising
background image

13

E n v i r o n m e n t a l c o n d i t i o n s a c c o r d i n g t o I E C 6 1 0 1 0 - 1 :

 Use only indoor.
 Altitude up to 2000 m - normal zero.
 Ambient temperature: see Technical specifications
 Air

humidity:

Max. rel. humidity 80 % for temperatures up to +31 °C,
linear decrease down to 50 % relative humidity at a temperature of +40 °C

 Max. mains fluctuations of ±10 % are permissible.
 Overvoltage category II
 Pollution degree 2

Caution:

The unit is not

suitable

for use in explosive environment

Protection class according to IEC 60 529: IP20
The unit corresponds to Class I

Standards for interference resistance according to EN 61326-1
This unit is an ISM device classified in Group 1 (using high frequency for internal purposes) Class A
(industrial and commercial range).

Advertising
This manual is related to the following products: