7.3. Warning functions

)) )) ))

More protection for your samples in the bath!
An audible signal sounds in intervals when the actual temperature value
exceeds one of the set limits (patented).

Overtmp:305.00°C
ExtAct: xxxxxx°C
Power: x %
Control: xxxxxx

Overtmp:240.00°C
ExtAct: xxxxxx°C
Power: x %
Control: xxxxxx

Setting the high temperature limit:

1.

Press the key

.

The value previously set appeears on the DIALOG-DISPLAY (LCD)
(example: 305.00°C). A flashing segment indicates that a value
needs to be entered.

2.

Use the keypad to enter the new value
(example: 240.00 °C).

3.

Press enter

to store the value.


Subtemp:220.00°C
ExtAct: xxxxxx°C
Power: x %
Control: xxxxxx

Setting the low temperature limit:

1.

Press the key

.

2.

Follow the instructions

3.

for

(example: 220.00 °C).

Note:

The warning functions will only be triggered when the actual bath
temperature, after start from the „OFF“ or „rOFF“ mode, lies within the set
limits for 3 seconds.