6. Operating

procedures

6.1. Power

connection

Caution:

 Only connect the unit to a power socket with an earthing contact (PE – protective

earth)!

 The power supply plug serves as a safe disconnecting device from the line and

must always be easily accessible.

 Never operate the unit with a damaged mains power cable.

 Regularly check the mains power cables for damage.
 We disclaim all liability for damage caused by incorrect line voltages!



Make sure that the line voltage and frequency match the supply voltage
specified on the type plate.
Deviations of ±10 % are permissible.

6.2. Switching on / Start - Stop

XXXXX

95xxxxx
-xxPxHx

-OFF-

S 150.0
I 24.6

Switching on:

 Turn on the mains power switch (1).

During the self-test all segments of the VFD-Info-Display light up.
Then the software version number (example: V 4x.1x) and the order
number of the recirculating cooler appears.
(Example: [ 95 20 025 03 P3H1 ] ).

The display „OFF“ or „R OFF“ indicates the unit is ready to operate.


The recirculating cooler enters the operating mode activated before
switching the recirculating cooler off:
keypad control mode

(manual operation)

or
remote control mode

(operation via personal computer).


Start:

Press the start/stop key

.

The actual bath temperature is displayed on the VFD COMFORT-
DISPLAY.

Stop:

Press the start/stop key

.

The VFD COMFORT-DISPLAY indicates the message „OFF“.

Autostart

: see chapter 9.3.2.

A-START – Autostart

The Autostart function enables the start of the recirculating cooler
directly by pressing the mains switch or using a timer.