6 Technical data (extract from data sheet)

30

Resistance transmitter and potentiometer

Input for DC voltage, DC current

Cu 100

GOST 6651-94 A.4

(TC = 4.26*10

-3

1/°C)

2/3-wire
2/3-wire
4-wire
4-wire

-50 to +100°C
-50 to +200°C
-50 to +100°C
-50 to +200°C

±0.5°C
±0.9°C
±0.5°C
±0.6°C

250μA

250μA

250μA

250μA

Connection circuit

2-, 3-, or 4-wire circuit

Shortest span

15°C

Sensor lead resistance

max. 30 per conductor for 3-wire/4-wire circuit

max. 10 per conductor for 2-wire circuit

Range start/end

freely programmable within the limits, in 0.1°C steps

Sampling cycle

Channel 1 - 18: 125ms in total

Input filter

2nd order digital filter; filter constant adjustable from 0 to 10 sec

Electrical isolation

see “Electrical data” on page 5 and

“Overview of the electrical isolation” on page 16

Resolution

>14 bit

Features

also programmable in °F

1

The linearization accuracy refers to the maximum measuring range. The linearization accuracy is reduced with short spans.

Designation

Meas. range

Accuracy

1

Meas. curr.

Resistance transmitter

up to 4000

Ω

±4

Ω

100μA

Potentiometer

< 400Ω

≥ 400Ω to 4000Ω

±400m

Ω

±4

Ω

250μA

100μA

Connection circuit

resistance transmitter: 3-wire circuit

potentiometer: 2-/3-/4-wire circuit

Shortest span

60

Ω

Sensor lead resistance

max. 30 per conductor for 4-wire circuit

max. 10 per conductor for 2-/3-wire circuit

Resistance values

freely programmable within the limits, in 0.1 steps

Sampling cycle

Channel 1 - 18: 125ms in total

Input filter

2nd order digital filter; filter constant adjustable from 0 to 10.0 sec

Electrical isolation

see “Electrical data” on page 5 and

“Overview of the electrical isolation” on page 16

Resolution

>14 bit

1

The linearization accuracy refers to the maximum measuring range. The linearization accuracy is reduced with short spans.

Basic range

Accuracy

1

Input resistance

-12 to +112mV
-10 to +210mV

-1.5 to

+11.5V

-0.12 to

+1.12V

-1.2 to

+1.2V

-11.2 to

+11.2V

±100

μV

±240

μV

±6mV
±1mV
±2mV

±12mV

R

E

≥ 1 MΩ

R

E

≥ 470 kΩ

R

E

≥ 470 kΩ

R

E

≥ 470 kΩ

R

E

≥ 470 kΩ

R

E

≥ 470 kΩ

Shortest span

5mV

Range start/end

freely programmable within the limits in 0.01 mV steps

-1.3 to

+22mA

-22 to

+22mA

±20

μA

±44

μA

burden voltage

≤ 3V

burden voltage

≤ 3V

Shortest span

0.5mA

Range start/end

freely programmable within the limits in 0.01 mA steps

Overrange/underrange

according to NAMUR NE 43

Sampling cycle

Channel 1 - 18: 125ms in total

Input filter

2nd order digital filter; filter constant adjustable from 0 to 10.0 sec

Electrical isolation

see “Electrical data” on page 5 and

“Overview of the electrical isolation” on page 16

Resolution

>14 bit

1

The linearization accuracy refers to the maximum measuring range. The linearization accuracy is reduced with short spans.

Designation

Standard

Connection circuit Meas. range

Accuracy

1

Meas. curr.

Advertising