Setup program (for all types)

The setup program is available for calibrating/configuring the transmitter from a PC.
Connection is through a USB/SPI-interface (including adapter) and the setup interface of the transmitter. In order to calibrate/configure
the transmitter, it has to be connected to the supply voltage. If no power supply or supply isolator is available, Types 707030/..., 707031/
... and 707032/... can be supplied from a 9V block battery.

Adjustable/configurable parameters

-

TAG number (8 characters)

-

response to probe and cable break

-

range start, range end

-

lead resistance for 2-wire circuit

-

measuring range configuration in °C or °F

Fine calibration

Fine calibration means adjustment of the output signal of a calibrated/configured transmitter. Errors due to the system (such as an
unfavorable probe installation) can be compensated. The signal can be adjusted in the range ±0.2mA for current output and ±0.1V for
voltage output. Negative output voltages are not possible with voltage output. Fine calibration can only be carried out through the setup
program.

Hardware and software requirements

The following hardware and software requirements have to be met for installing and operating the setup program:
-

IBM-PC or compatible PC

-

256 MB main memory

-

50 MB available on hard disk

-

CD-ROM drive

-

1 USB interface

-

Windows 2000, XP, Vista, Windows 7 (32 Bit and 64 Bit)

Connection layout for calibrating/configuring the dTRANS T03 B and BU

Calibration
(basic type extension 880/990)

Configuration
(basic type extension 881/991)

USB/SPI

+

-

+

-

-

+

Supply voltage

7.5 — 30V

/ 25mA

DC

Indicator

4 — 20mA

Precision
decade resistor

Supply voltage

15 — 30V DC

Indicator

0 — 10V

Precision
decade resistor

dTRANS T03 BU

dTRANS T03 B

USB

USB/SPI converter

Adapter

USB/SPI

+

-

+

-

Supply voltage

7.5 — 30V

/ 25mA

DC

Supply voltage

15 — 30V DC

dTRANS T03 BU

dTRANS T03 B

USB

USB/SPI converter

Adapter