Technical data

Analog input

Measuring input

Pt1000 according to DIN EN 60751, in 3-wire circuit

Operating temperature range

-30 to +260 °C

Accuracy
of the temperature sensing ele-
ment

Class A ±(0.15 K + 0.002 × ItI)
ItI = measured temperature in °C without prefix

Sensor lead resistance

11 ohm per lead with 3-wire circuit

Sensor current

500 µA

Measuring circuit monitoring

Detection of probe break and probe short-circuit

Output (radio transmission)

Transmitter detection
(transmitter ID)

Max. 5-digit ID, factory-set, cannot be configured

Transmission interval

Adjustable via DIP switch to 5 s, 10 s, 20 s, or 45 s
(factory-setting = 20 s)

Radio frequency

868.4 MHz

Transmission power

< +10 dBm

Open air range

300 m and use of the receiver
antenna wall holder, and 3 m antenna cable.
When installing the antenna directly on the receiver, the user must
take approx. 40 % less range into consideration.

Output signal

882.2 to 1977.1 ohm = -30 to +260 °C

Response time
of the complete probe

In water 0.4 m/s / in air 3.0 m/s

4.0 mm:

Water t

0.5

approx. 3 s, t

0.9

approx. 7 s / air t

0.5

approx. 25 s, t

0.9

approx. 80 s

6.0 mm:

Water t

0.5

approx. 4 s, t

0.9

approx. 10 s / air t

0.5

approx. 32 s, t

0.9

approx. 98 s

Calibration accuracy
of the electronic components

±0.05 %

1

1

All accuracy specifications in % relating to the measuring range of 290 K.

Voltage supply

Lithium battery

Voltage: 3.6 V, rated capacity: 2.2 Ah

Operating life

Approx. one year with the factory-set values and at room tempera-
ture (a fast transmission interval and a high or low ambient temper-
ature reduce the battery operating life)

Battery change

Only use the preconfigured lithium battery that is available in the
accessories (sales no. 90/00525539)