11 fault finding, 1 what to do if, Fault finding – JUMO 706021 LOGOLINE 500d Line Recorder Operating Manual User Manual

Page 119: What to do if

Advertising
background image

119

11 Fault finding

11.1 What to do if...

Display or printing

“>>>>>>”

-

The value is outside the measuring range (Out Of Range)

-

The transducer has been connected up incorrectly

-

The signal inputs are configured incorrectly

-

Probe break

Instead of a value, the prin-
ting or the display shows

“****”, “±*.***”, “±**.**”,
“±***.*”, “±****.”

-

The value can not longer be represented; alter the scaling with the
setup program so that the value can always be represented.

-

In the case of event counter, reset the event counter using the setup
program

No recording

-

Key

S

has been pressed

 Section 6.1

-

“External Stop” has been configured and the corresponding logic in
put is closed

-

The chart cassette has not been fitted correctly
 Section 9.2
Caution: Insert the chart cassette by applying light upwards pressure

-

The take-up roll has not properly engaged with the chart cassette
 Section 9.2.1

-

The end of the chart has been reached
 Section 9.2.1, Section 9.2.2

-

The chart speed has been programmed as 0 mm/h
 Section 8.2.1

-

The writing status is switched off (“OFF”)

-

The setup connector is still plugged in

Fibre pen does not write

-

The fibre pen has not been fitted correctly
 Section 5.4

-

The ink in the fibre pen is used up

No chart feed

-

The chart cassette has not been fitted correctly
 Section 9.2

-

The chart speed has been programmed as 0 mm/h

-

The pin roller does not engage with the chart perforations

-

The tear-off edge is not correctly engaged

-

The pen recorder is in the STOP status

-

With a newly inserted chart, the start of the chart must be straight and
smooth so that the chart can be taken up automatically

Measurements are not
recorded

-

Check that the connecting terminals are tightened properly

-

Check the supply

-

Check the input configuration (range)

-

Check the transducers and cables, measure them where appropriate

-

The write status may be switched off

Advertising