Kds model 410 series syringe pumps, Operating instructions – KD Scientific 410 User Manual

Page 18

Advertising
background image

16

KDS Model 410 Series Syringe Pumps

KDS Model 410 Series Syringe Pumps

Operating Instructions

(Continued)

RS232 COMMANDS (continued)

Prompts

>

running in infusion direction

<

withdrawing

:

stopped

NA

not applicable

E

error (see error? command)

carriage return <CR>

All pumps chain interpret this as a stop command

pump address<CR>

Pump with the specified address responds with a prompt

pump address (optional), Pump at the address executes the command and then

command<CR>

responds with a prompt.

Note: If there are multiple pumps in the daisy chain and a pump address is not used

then all pumps will respond to the non-specific command and return prompts. Multiple

prompts results in a communications breakdown.

Pump commands and responses

Note that mode selection and withdrawal and continuous mode commands are recognized

only by the infusion/withdrawal models.

run

Starts pump running to present settings, returns prompt > or < If already

running command is ignored.

stop

Stops pump if running, otherwise is ignored. Returns prompt :

dia nn.nn

Sets syringe diameter in millimeters.

n = 0 to 9

run?

Query run status, returns prompt.

del?

Queries delivered volume. (Requires a dispense volume to be set).

Response: nnnnn<SP>u

Where nnnnn is ., 0 to 9

u are units, µl or ml

dia?

Requests present diameter setting.

Response: nn.nn

Where n = 0 to 9

units always millimeters

ratei?

Queries infusion rate.

Response: nnnnn<SP>u/u

Where nnnnn is ., 0 to 9

u/u are units, µl/m, µl/h, ml/m, or ml/h

ratew?

Queries withdrawal rate.

Response: nnnnn<SP>u/u

Where nnnnn is ., 0 t o 9

u/u is rate unit