Kds milliliter oem module, I/o specifications: inputs, Outputs – KD Scientific 910 OEM Pump User Manual

Page 15

Advertising
background image

16

KDS Milliliter OEM Module

16

The pump is set to the following default parameters on power-up and after an

external enable command.

DEFAULT SETTINGS

Syringe Diameter:

2.300 mm

Rate:

3.000 ul/min

Range:

ul/min

Force:

100%

Address:

00

Baud Rate:

9600

I/O Specifications: Inputs

RUN/STOP

This is an active low, TTL level input, pulled up to +5V through a 10K ohm resistor. It is ESD

protected through a TVS device and filtered with a 0.1uF capacitor to ground. Each pulse to

a logic low toggles the pump between the RUN and STOP states.

EXT_ENABL/

This is an active low, TTL level input, pulled up to +5V through a 10K ohm resistor. It is ESD

protected through a TVS device and filtered with a 0.1uF capacitor to ground. A transition

from logic high to logic low causes the processor to reset to its default state and enables the

motor drive. A transition from logic low to logic high disables the motor drive through hard-

ware (independent of firmware). This input may be used for an emergency stop function.

**To enable pump, the unit or connector P7 pin 3, must be connected to ground**

Outputs

RUN_IND/

This is an active low output driven by two 74HCT14 inverters in parallel. An on-board resis-

tor may be placed in series with this output to provide current limiting. The default resistor

value is zero ohms. A logic low indicates RUN. A logic high indicates STOP.

INDICATORS:

Power-on LED

When illuminated, indicates that board is powered on and +5VDC supply is operating.

RUN LED

When illuminated, indicates that pump is running. When extinguished, indicates pump is

stopped.