29

KDS Gemini 88

Appendix H: Troubleshooting

RS-232 DIFFICULTIES

Verify that the baud rates and data framing parameters on all devices are the same. With a

pump chain, a “Communication timeout” error on a computer is usually caused by the com-

puter errantly handshaking on the RTS, CS and/or DSR lines. Verify pins 4, 5 and 6, on the

25 pin connector, are jumpered on the computer side of the cable.

DISPLAY MESSAGES:

All digits flashing

Pump has not received any commands since being powered on.

Stall

Also a syringe LED and Run LED flashing. This indicates that the pusher block travel of the

syringe corresponding to the syringe LED blinking has been impeded. This may be caused

by the syringe plunger hitting bottom, a kink in the tubing, syringe plunger binding or any sit-

uation requiring more force to the head of the syringe than the pump is capable of deliver-

ing. To restart the pump, remove the cause of stalling and press RUN. The pump will contin-

ue the interrupted procedure where it stopped. Also note, a large increase in the pumping

rate while running could also stall the motor.

oor

Out of range. A value was entered or encountered in a pump program that was beyond the

pumps limits. Try using a different size syringe. Note that rates must be less than 42950.

Express larger rates using different units.

- - - - -

Key pressed is not applicable.

Appendix I: Accessories

Catalog No.

Product

MA1-55-4144

Foot Switch

MA1-55-7760

Daisy Chain Cable 2ft for connecting two or more pumps; need 1

cable per pump

MA1-70-2022

RS-232 Connector Cable, 7ft, 9 pin D-Sub

MA1-55-7000

Continuous Delivery Valve Box, normal pressure, 30 p.s.i. supplied

with 1/8” ID x 1/4” OD silastic tubing and connector cable for

connection to the syringe pump.

MA1-55-7001

Continuous Delivery High Pressure Valve Box, high pressure, 200 p.s.i.

supplied with 1/4” OD 304 stainless steel tubing with SWAGELOK

®

fittings and a connector cable for connection to the syringe pump.