0 technical specifications – Labnet 222DS Benchtop Shaking Incubator User Manual User Manual

Page 10

Advertising
background image

BENCHTOP SHAKING INCUBATOR 222DS

USER'S MANUAL

Page 10

5.2

DESCRIPTION OF POSSIBLE ERRORS


E 11:

This error message appears, when something is wrong with motor regulation (PWM
regulator, pulse generator, motor). Shaking incubator is automatically stopped. Call
nearest service.

E 12:

This error appears, when the motor does not reach set speed in 30 seconds.
Shaking incubator is automatically stopped. Call nearest service.

E 13:

This error appears, when set speed oscillates for more than 100 RPM in 2 seconds.
Shaking incubator is automatically stopped. Call nearest service.

E 21:

Temperature sensor does not work properly regarding to heater. Shaking incubator
is automatically stopped. Call nearest service.

E 22:

Temperature sensor notices 5°C higher temperature than set temperature. Shaking
incubator is automatically stopped. Call nearest service.

E 23:

Set temperature is not reached in 2 hours. Shaking incubator is automatically
stopped. Call nearest service.

6.0 TECHNICAL SPECIFICATIONS


6.1

TECHNICAL DATA


Heater power

700 W

Fan power

17 W

Shaking motor power

35 W

Fuse

2 x 3.15A 250V
2 x 6.3A 115V

Environment temperature

4°C to 65°C

Relative humidity

Up to 85%, non condensing

Speed RPM regulation

Digital, load independent,
from 20 to 300 RPM in 1 RPM steps

Shaking orbit

19 mm

Temperature operating range

5°C above room temperature to 70.0°C

Temperature sensor

PT100

Temperature accuracy

±0.5°C

Temperature stability

±0.5°C

Timer

1 min - 99 hours, timer HOLD function

Capacity

16 x 125 ml, 9 x 250 ml, 5 x 500 ml, 4 x 1000 ml

Dimensions W x D x H

370 mm x 530 mm x 400 mm

Weight 20.2

kg

Overvoltage category

Installation category I