3


Rotor

Lid

(Included)

Order

no.

C0240-RC


Rotor

Securing

Screw

(Included)

Order

no.

C0240-RSS

Strip Spin Adapter

Order

no.

C0240-SS

Tubes accepted

0.2ml tubes or 8 x 0.2ml strips

Max

Speed

13,300

rpm

Range Centrifuge radius

4.32cm to 5.33cm

Range RCF (g value)

8,549 x g to 10,539 x g


Adapter for 0.5ml tubes (pk./6)

Order

no.

C-1205

Tube

measurement

8

x

30mm

Max.

speed

13,300rpm

Centrifuging

radius

7.53cm

RCF

(g-value)

14,892

x

g


Adapter for 0.4ml tubes (pk./6)

Order

no.

C-1206

Tube

measurement

6

x

47mm

Max.

speed

13,300rpm

Centrifuging

radius

8.23cm

RCF

(g-value)

16,276

x

g


Adapter for 0.2ml tubes (pk./6)

Order

no.

C-1222

Tube

measurement

6

x

21mm

Max.

speed

13,300rpm

Centrifuging

radius

7.03cm

RCF

(g-value)

13,903

x

g


3.2 Rotor maintenance

The rotor should be cleaned thoroughly after each use. Thorough cleaning must be

performed when spinning samples containing phenol or phenol chloroform.

Periodically inspect the rotor for dents, dings, scratches, discoloration and cracks. If any

damage to the rotor is found, discontinue use of the rotor immediately and replace.


3.3 Removing and installing the angle rotor

The Spectrafuge 24D comes complete with a standard 24 place rotor installed. To remove

the rotor for cleaning, remove the rotor securing screw from the motor shaft by turning

the screw counterclockwise, using the rotor wrench. Lift the rotor directly upward in a

straight vertical motion. To replace rotor, first make sure the motor shaft and rotor

mounting hole are clean. Place the rotor on the motor shaft. Reinstall the rotor securing

screw on the motor shaft by turning it clockwise. Hold the rotor with one hand and

tighten the rotor securing screw, using the rotor wrench.