Safety Precautions:

NEVER use the centrifuge in any manner not specified in these instructions.

NEVER

put hands in the rotor area unless the rotor is completely stopped.

NEVER

move the centrifuge while the rotor is spinning.

NEVER

use solvents or flammables near this or other electrical equipment.

NEVER

centrifuge flammable, explosive or corrosive materials

NEVER

centrifuge hazardous materials outside of a hood or proper contain

ment

facility

NEVER

continue to operate a centrifuge that is vibrating excessively.

ALWAYS load the rotor symmetrically. Each plate should be counterbalanced by

another plate of the same brand, type and volume.

ALWAYS use plates that have been sealed in some manner to avoid spillage.

ALWAYS load plates so that the sealed side faces the center of the rotor and the

bottom of the plate faces out.

ALWAYS be sure the plates are pushed all the way to the bottom of the rotor slot

before operating the centrifuge.

ALWAYS locate the centrifuge within easy access to an electrical outlet.

Cleaning:

Before using any cleaning or decontamination methods except those recommended by
the manufacturer, check with the manufacturer that the proposed method will not
damage the equipment. To clean the centrifuge, use a damp cloth and a mild,
noncorrosive detergent (ph <8). After cleaning, ensure all parts are dried thoroughly
before attempting to operate the unit. Do not immerse the centrifuge in liquid or pour
liquids over it.


Questions and Service:

Should you have a question about the centrifuge or require service for the unit, please
use the contact numbers listed below. Do not send in a unit for service without first
calling to obtain a repair authorization number and a decontamination form. Should the
centrifuge require service, be sure to decontaminate it and enclose the decontamination
form. The unit should be properly packed to avoid damage. Any damage resulting from
improper packaging shall be the responsibility of the user.


Specifications:

Dimensions... ........ 6” x 6” (153 x 153mm)

Electrical ............... 115V, 50/60Hz,130mA

Fuse ..................... 0.5A, 250V (F)

Speed ................... 6000rpm

RCF ...................... 2000 x g Ambient Operating Range: 0 to 40°C,< 85%