Labnet Z400 High Capacity Centrifuge User Manual

Page 13

Advertising
background image

4.4

Preselection of speed / RCF

You can preset the speed between 250 rpm amd 13500 rpm in steps of
250 rpm

4

1

5

3

2

When the centrifuge lid is open, you can preset the required speed or rcf with the
knob(2).
If the centrifuge lid is closed and during the run, the speed can be changed as
follows:
Press key "preset" (5), hold it and at the same time change speed with the knob
(2).
The preselected speed will be indicated in the speed display (4).
The centrifuge will accelerate only to the max. allowed speed of the inserted rotor.
The allowed max. speed for all rotors please see table under.

Z 400

Z 400

Z 400

Z 400

Z 400

Z 300

Z 300

Z 300

Z 300

Z 300

Rotor

max. Speed

Rotor

max. Speed

220.86 V01

3 500 rpm

220.72 V04

4 000 rpm

220.27 V02

4 000 rpm *

220.50 V05

3 500 rpm

220.50 V05

3 500 rpm

220.81 V05

3 500 rpm

220.81 V05

3 500 rpm

220.82 V01

6 000 rpm

220.82 V01

6 000 rpm

220.59 V06

13 500 rpm

220.59 V06

13 500 rpm

*with bucket 611.015 3500 rpm