Instrument setup – Leica Biosystems SM2010 R User Manual

Page 14

Advertising
background image

12

Instructions for Use V 1.4 RevC – 09/2013

4.

Instrument Setup

25

26

Horizontal alignment
For safe and accurate work, it is important that

all instrument feet are in uniform contact with the

installation surfaces.
The microtome is horizontally aligned at the fac-

tory. If a completely level or horizontal surface is

not available at the installation site, the instrument

must be realigned.
To do so, the two instrument feet (25) on the right

side of the instrument are height-adjustable.

• For alignment, loosen the locknuts (26) using

a size 10 open-end wrench.

• Adjust the instrument feet (25) until the micro-

tome is in a stable position at the installation

location according to requirements.

• Retighten the locknuts.

Setting the Ergo handle
The Ergo handle (12), which is used to move the

knife sledge, can be set individually to an ergo-

nomic gripping position.
• To do this, loosen the fastening screw (10)

using a size 6 Allen key (28).

• Turn the handle (12) to the desired position

and retighten the screw (10).

12

28

10

Fig. 7

Fig. 8

Advertising