Leica VT1000 S - Vibrating-blade microtome

6. Malfunctions: Meaning and Troubleshooting

Error messages/symptoms

Sources of error

Troubleshooting

- Audible warning signal.

- Downward stroke is not com-

pleted.

- Operating error due to locking

function of the UP/DOWN button:

- With the UP/DOWN button

locked in the DOWN position

the instrument was switched off

via the power switch at the rear

of the instrument switched on

again without releasing the UP/

DOWN button to its center posi-

tion.

- Release the UP/DOWN button to

its center position.

- To reactivate the downward

motion, activate the UP/DOWN

button again (DOWN).

- Audible warning signal.

- Downward stroke is not com-

pleted.

- With the UP/DOWN button

locked the instrument was

switched off via the EMERGEN-

CY STOP (foot switch or Emer-

gency stop button) and after

that the EMERGENCY STOP was

released without unlocking the

UP/DOWN button.

- Release the UP/DOWN button to

its center position.

- To reactivate the downward

motion, activate the UP/DOWN

button again (DOWN).

- The feed motor stops.

- Any processing step (sectioning

stroke etc.) is interrupted imme-

diately.

- Any UP/DOWN motion of the

buffer tray is interrupted imme-

diately.

- Any locked buttons are indicat-

ed by an audible warning signal.

- When pressing any key, the

instrument gives an audible

warning signal.

- In case the EMERGENCY STOP

function has been activated, the

instrument will remain inopera-

tional when pressing the foot

switch.

- The indication SP is displayed.

- The EMERGENCY STOP function

has been activated.

- Release the Emergency stop

button.

- Select an operating mode and

continue working.