Cleaning and maintenance, Cleaning the instrument, Replacing fuses – Leica Biosystems HI1220 User Manual

Page 14

Advertising
background image

14

Instructions for Use V 3.2 RevE – 02/2013

6.

Cleaning and Maintenance

6.1 Cleaning the instrument

• Before cleaning the instrument, turn off using the power switch (

Fig. 6

, 3) and disconnect

from the power supply.

• The instrument has a continuous front panel and no lateral cooling slots, so the surfaces

are smooth and easy to clean.

• All surface materials are resistant to common laboratory detergents.

The instrument must be switched off and have cooled down for cleaning.

Do not use organic solvents (e.g. alcohol or xylene)!

6.2 Replacing fuses

Switch the unit off and unplug it. Use

only the supplied replacement fuses.

Both fuses must have the same rating

(check the imprint!). Fuses must always

be switched out in pairs.

Fig. 8

• Insert a small screwdriver into the cutout

(

Fig. 8

, 11) and carefully pry out the fuse

insert (12).

• Remove the fuse holder and replace the de-

fective fuses with two new ones (13).

• Then reinsert the fuse holder and press it

into place.

12

13

11

х