Leica CM1510 S – Cryostat

Before replacing the lamp, turn off
instrument mains switch and pull
mains plug!

Hold sliding window (1) at grip (2), lift slightly
and pull it out to the front.

If the lamp is broken, it must be re-
placed by the technical service -
high risk of injury!

Removing the lamp

Lamp (3) is mounted behind glare shield (4)
and therefore not visible.

Touch the lamp for better orientation.

Lightly tilt the fluorescent tube down to the
left and pull it out of clip (5).

Hold lamp with both hands and pull it to the
left, out of socket (6).

Do not use any lamp type other than
the one inserted by the manufac-
turer!

8.

Cleaning and maintenance

5

4

3

6

Fig. 33

Fig. 34

Installing the new lamp

Lamp type:
(230 V/ 50 Hz) OSRAM DULUX S - 11 W
(115 V/60 Hz) OSRAM DULUX S - 13 W

Hold lamp in correct mounting position -
see Fig. 34 - and push to the right until it en-
gages in the socket.

Lightly push the lamp upward to engage in
the clip.

Reinsert sliding window.

Reconnect instrument to mains power and
turn it back on.

8.8

Replacing the lamp